1200°C Muffle Furnace (12x12x12", 27 L) with Programmable Controller & Venting Port
SPECIFICATIONS
TOB-MF1200Muffle furnaceis a 1200℃ laboratory muffle furnace mainly used for battery materials or other powder materials and ceramic sintering research.
Model
|
Muffle Furnace TOB-MF1200
|
Structure
|
1 Thermal efficient ceramic (high purity alumina fiber) insulation surrounding
the chamber provides maximum energy efficiency.
2 Heating elements on all four sides of interior chamber for fast heating and
enhanced temperature uniformity.
3 Heavy duty double layer structure with cooling fan to keep furnace exteriors safe to touch.
4 Built-in vent port removes undesirable contaminants and moisture;
extends the life span of the heating elements and furnace.
|
Chamber Dimensions
|
Size: 12" x 12" x 12", 300*300*300mm
Capacity: 27 liters
|
Power
|
7 KW Max. at 10oC/min to 1200oC
40 A fuse (installed)
|
Voltage
|
110~240V AC, Single phase, 50/60Hz
|
Operating Temperature
|
1100°C (Continuous)
1200°C Max (< 3 hour)
|
Temperature Stability
|
±1°C at 1000°C
|
Temperature Uniformity
|
±3.45°C at 1000°C
|
Max. Heating Rate
|
10°C/min
|
Temperature Controller
|
1 PID automatic control with 30 programmable segments for precise control.
2 Built-in over temperature alarm and protection
3 +/- 1 ºC temperature accuracy.
4 K type thermocouple installed for immediate use.
|
Gas Fitting & Exhaust
|
One 1/4" barbed Gas Inlet installed on back panel of furnace
One gas exhaust port is built on the top of the furnace casing.
|
Product Dimensions
|
24" x 26" x 22" (D x H x W) 60*66*55cm
|
Shipping Dimensions
|
48" x 40" x 49" (D x W x H) 122*100*125cm
|
Shipping Weight
|
350 lb, 160kg
|
Warranty
|
One year limited warranty with life time support.
|
Application Notes
|
1 For preserving the longevity of the furnace components,
we recommend the heating rate not to exceed 10 °C /min.
Cooling rate shall also not exceed 10 °C /min.
2 This furnace lacks vacuum sealing ,
toxic or explosive gases are not recommended for use with
this furnace without necessary safety controls and supervision.
3 Small cracks may appear on surface of refractory ceramics over extended use.
This is a normal occurrence and the cracks may be repaired with alumina coating
|
Email : tob.amy@tobmachine.com
Skype : amywangbest86
Whatsapp/Phone number : +86 181 2071 5609