Conductivity Meter With Temperature Compensation Function
SPECIFICATIONS
This TOB-DDS-307Aelectrical conductivity metercan testConductivity, TDS, temperature, selecting electrode measured high pure water.
Model
|
Conductivity Meter With Temperature Compensation TOB-DDS-307A
|
Source voltage
|
AC 220V/110V, 50/60HZ
|
Warranty
|
One year limited warranty with lifetime support
|
Advantages
|
1 Big screen, blue backlight, double digital display LCD
2 With manual, automatic temperature compensation function
3 Electrical connector below the plane, helps to corrosion protection and security
|
Testing range
|
Electrical conductivity : (0~1×105)μS/cm
TDS : (0~1000) mg/L
Temperature :(0~60)℃
|
Testing error
|
Electrical conductivity : ±0.5%(FS)±1
Temperature : ±0.3℃
|
Stability
|
±0.33%(FS)±1/3h
|
Constant conductance pool with
best conductivity measurement range
|
Electrode constant
(cm-1)
|
0.01
|
0.1
|
1
|
10
|
Test range
(μS/cm)
|
0~2
|
0.2~20
|
2~1×104
|
1×104~1×105
|
|
Temperature compensation scope
|
(0~35))℃
|
Dimension(L*W*H)
|
300×200×72mm
|
Weight
|
About 2kg
|
Email : tob.amy@tobmachine.com