Horizontal Vertical Flame Test Chamber ZLT-HV500
Model: ZLT-HV500
The ZLT Horizontal Vertical Flame Chamber Tester is a fully automatic instrument contained in its own cabinet to maximize the safety of the operator. It screening procedure for comparing the relative burning behaviour specimens made from plastic and other non-metallic materials ,exposed to a flame ignition source of 500W nominal power.
These test methods determine the linear burning rate and the after flame/after glow times, as well as the damaged length of specimens. They are applicable to solid and cellular materials that have an apparent density of not less than 250kg/m3,determined in accordance with ISO845.Then do not apply to materials that shrink away from the applied flame without ignition;ISO9773 should be used for thin flexible materials. And it's the combustion test device of plastic materials represented by UL 94. It can perform each test of HB ,5VA V5VB .
Specification:
Model |
ZLT-HV500 |
Test Flame |
500W (burning test for 5VA , V5VB classification ) |
Burner Tube |
Internal Diameter: Φ 9.5 mm ±0.1mm, length : 110mm ±0.1mm |
Burning Angle |
0“, 20” , 45" |
Thermocouple |
Type K ( Ni/ Cr -Ni/Al ) |
Thermocouple Size O/D |
0.5 mm |
Standard Copper Block |
Φ 9mm ± 0.01 mm, weight: 10g ±0.05g before drilling Cu-ETP |
Time for Temperature to increase |
From 100°C ±5°C to 700°C ±3°C: 54s ± 2s (preset) |
Overall Height of flame |
38mm - 42mm (adjustable ) |
Gas Supply (do not provide ) |
Methane min.98% purity |
Flowmeter and Manometer |
965 ± 30ml /min. 0 Kpa to 7.5 Kpa |
Input Power |
220V 50Hz 3A or 115V 60Hz, optical |
Coform to Standard |
UL94, IEC 60695-11-3, IEC 60695-11-20 |
Include: 500W Flame Gauge and 100% Absorbent Cotton .