Lab 1200C Single ZoneTube Furnace with Many Tube Size Optional
This tube furnace uses resistance wire as heating element, double-layer shell
structure and 30-stage program temperature control system, phase shift
triggering, thyristor control, and furnace polycrystalline fiber material. It
has the advantages of temperature field balance, low surface temperature, fast
temperature rise and fall, energy saving, etc. It is an ideal product for
high-temperature atmosphere sintering, atmosphere reduction, CVD experiment and
vacuum annealing in universities, research institutes and industrial and mining
enterprises.
Detailed technical parameters
Type
|
1200-I
|
1200-II
|
1200-III
|
1200-V
|
1200-VI
|
1200-VII
|
Power
|
3.5KW
|
3.5KW
|
3.5KW
|
5KW
|
6KW
|
7KW
|
Tube size(OD(mm)
|
Dia 60 X1000
|
Dia80X1000
|
Dia100X1000
|
Dia150X1000
|
Dia200X1000
|
Dia250X1000
|
Dimensions
|
1140X450X690
|
1140X450X690
|
1140X450X690
|
1110X530X755
|
1120X580X805
|
1314X680X905
|
Supply voltage
|
220V
|
Phase
|
Single phase
|
Heating element
|
Resistance wire
|
Temperature control
|
UAV program temperature control
instrument (standard)
1, 30-stage program temperature
control intelligent PID adjustment.
2. With over-temperature protection,
the electric furnace heating circuit is automatically cut off when the
temperature is over-temperature or broken, (when the electric furnace
temperature exceeds 1200 degrees or the thermocouple is blown, the AC relay
on the main circuit will be automatically disconnected, the main circuit is
broken. On, the ON light on the panel is off, the OFF light is on, and the
limited protection electric furnace).
3, with 485 communication interface
(standard when purchasing software)
4, with power-off protection
function, that is, when the power is turned on after the power is turned off,
the program does not start from the starting temperature, but the furnace
temperature rises from the time of power failure.
5, the
meter has the function of temperature self-tuning
(you need to pay extra for
the replacement of imported instruments, optional)
FP93 temperature controller (shimashima,
Japan) EUROPEAN instrument Touch screen instrument
|
Chamber material
|
1.
High-quality high-purity alumina polycrystalline fiber curing furnace formed
by vacuum suction filtration.
2.
Formed by Japanese technology.
3. The spacing and pitch of the resistance
wires in the furnace are all arranged according to the best thermal
technology in Japan, and the temperature field is simulated by the thermal
software.
|
Valve and flange
|
|
60/80/100top view
|
|
Temperature calibration
(optional)
|
Regular
temperature calibration of the tube furnace maintains the accuracy and
uniformity of the furnace temperature. You can purchase a simple calibration
tool from the company to assist you with this work. The complete calibration
kit includes a standard thermocouple S-type, temperature calibrator.
|
Control accuracy
|
+/- 1℃
|
Trigger
|
Phase-shifted
trigger
|
Electric appliance
|
Zhejiang chint
|
Silicon controlled
|
106/16ESEMIKRON
|
Maximum temperature
|
1200℃
|
Rated temperature
|
1150℃
|
Heating rate
|
≤30℃/ Min(it
can be modified as required)
|
Recommended rate of
temperature rise
|
≤15 ℃/ Min
|
Thermocouple Type
|
K Type
|
Heating length
|
400-440mm
|
Constant temperature length
|
150mm
|
Temperature field curve
|
|
Peripheral surface temperature
|
≤45℃
|
Open cover protection
system
|
When the furnace cover is
opened, the relay will automatically cut off the main power supply,
effectively ensuring the safety of the user.
|
1 Standard exported package: Internal anticollision protection, external export wooden box packaging.
2 Shipping by express, by air, by sea according to customers' requirements to find the most suitable way.
3 Responsible for the damage during the shipping process, will change the damage part for you for free.