Safety Shoes Cold Insulation Test Chamber GT-KB25
Product Introduction
Application:
Safety Shoes Cold Insulation Test Chamber is used to determine the cold insulation test for safety shoes under low-temperature.
Standards:
ISO 20344 section 5.13, GB/T 20991 section 5.13, AS/NZS 2210.2 section 5.13
|
Dimensions
Feature
1.Adjust the temperature of the cold box to (–17 ± 2) ℃ and maintain this temperature during the test. 2.Place the test piece on the laboratory jack inside the cold box.
3.Use the temperature measuring device connected to the temperature probe to measure
the temperature on the insole as a function of time, recording the temperature decrease graphically. 4.Calculate, to the nearest 0,5°C, the decrease in temperature 30 min after the test piece was placed
in the cold box.
Key Specifications