XT 9160-T AZ14-1 KD Product Overview:
1.A common fault with a capacitor discharge mobile is a short circuit in the cathode cable, between the X-ray tube grid and cathode connections. This results in:
a) Although preparation is obtained, an exposure cannot be made.
b) The capacitor will not charge to the required kV.
c) As the generator is placed into preparation, an
immediate exposure takes place, discharging the capacitor.
2. The films from the processor are discoloured, and feel sticky.
a) The wash water temperature is too high.
b) Poor quality films. Try a new box.
c) Fixer is depleted or contaminated.
3. Developer temperature should be checked.
a) Weekly.
b) Twice a week.
c) Daily.