Walk-in Stability Test Chamber 15000SD

The XCH-15000SD Series newly airway system designed achieves uniform of temperature and humidity form different parts inside the chamber. The inner wall and duct plate are made of stainless steel 304. Observation window of hollow tempered glass with electrothermal film.


Temperature Range: 20~ 50℃

Temperature Fluctuation≤ ±0.5℃

Temperature Deviation≤ ±1.0℃

Humidity Range40~80%RH(or 2080%RH);

Humidity Deviation≤ ±3.0%RH

Test Points Optional40℃ /75%RH25℃ /60%RH30℃ /65%RH

40℃ /25%RH25℃ /40%RH25℃ /60%RH

Low Temperature Laboratory: 2 8℃



Walk-in Constant Temperature and Humidity Stability Chamber


Performance:

1.Reference Standard: Refer to GB/T 10586-2016, JJF1101-2019 environmental test equipment temperature and humidity calibration specification for technical conditions of hot and humid test chamber
2.Temperature Range: 20 ~ 45℃,Temperature Fluctuation:≤ ±0.5℃ ,Temperature Deviation:≤ ±1.0℃
3.Humidity Range:40 ~ 80%RH,Humidity Deviation:≤ ±3.0%RH(Customizable Humidity Range:20 ~ 80%RH)
4.Test Points Optional: 25℃ /60%RH、30℃ /65%RH( Low humidity : 40℃ /25%RH、25℃ /40%RH);
5. Low Temperature Laboratory: The cool storage can be customized under 20℃ and 2 ~ 8℃ cold storage (Temperature Fluctuation ≤ ±1.0℃,Temperature Deviation ≤ ±2.0℃ )

Key Features:

Structure System:  

The newly airway system designed achieves uniform of temperature and humidity form different parts inside the chamber. The inner wall and duct plate are made of stainless steel 304. Observation window of hollow tempered glass with electrothermal film. Illumination lamp, control panel with running indicator light and MIL malfunction indicator lamp.

Refrigerating System: 

Original imported refrigeration compressor unit with high performance、low noise and one work one back up.Control System:  Original imported programmable chromatic touch screen controller, sensitive, Small system error, Multi - stage program setting and multi - point accurate calibration function, With three level password;Humidity System:  The world’s top VAISALA original imported humidity transducer with high precise,Low drift,Long-life and maintenance-free.

Date System: 

Needle micro printer,SD card storage and online electronic recording.Alarm System:   Sound-light alarm for temperature and humidity deviation, SMS alarm for remote temperature and humidity deviation and power failure, email alarm.

Safety Device:

The door is equipped with key safety lock and anti-reverse lock function to ensure the safety of users. The door frame is equipped with unique anti-condensation and anti-leakage device, compressor overload protection,  water shortage protection, anti-dry burning protection system, independent over-temperature protection system.

Electronic recording system:

Comply with the FDA 21 CFR Part Ⅱ law standard, Stable system performance, Audit tracking function, Three levels of user hierarchical authority management.


Walk-in Stability Chamber Specification:

MODEL
TEMP RANGE(℃)
HUMIDITY RANGE
CAPACITY(L)

INTERNAL DIMEN(mm)

W×D×H

EXTERNAL DIMEN(mm)

W×D×H

POWER(KW) REMARKS
XCH-8000SD
20 ~ 45
40 ~ 80%
8000 1900×2200×2000
2100×3100×2300
5.0
Customizable Humidity: 20 ~ 80%RH 
XCH-15000SD
20 ~ 45
40 ~ 80%
15000 1900×4000×2000
2100×4900×2300
6.0
XCH-20000SD
20 ~ 45
40 ~ 80%
20000 2200×4100×2200
2400×4300×2400
6.5
XCH-40000SD
20 ~ 45
40 ~ 80%
40000 3600×5200×2200
3800×5400×2400
8.5


Advantages:

Electronic recording system

Comply with the FDA 21 CFR Part Ⅱ law standard

Stable system performance

Audit tracking functioncustomizable



Product Tags Walk-In Stability test Chamber          Best Walk-In Stability Chamber          Medicine Walk in stability chamber          Walk-In chamber         

INQUIRY

Contact us for all information about the product


click here to leave a message

leave a message
If you are interested in our products and want to know more details,please leave a message here,we will reply you as soon as we can.